TG1002
FUTEK reserves the right to modify its design and specifications without notice.
Please visit www.futek/salesterms  for complete terms and conditions.          10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900
www.futek
Important Tips . . . . . . . . . . . . . . . . . . . . . . 3Potential 5Programming Tips . . . . . . . . . . . . . . . . . . . . 9
.11Technical Support . . . . . . . . . . . . . . . . . . . .12Contact Information . . . . . . . . . . . . . . . . . . .13
Table of Contents
The USB Troubleshooting Guide can be found at: media.futek/docs/troubleshooting/
usb_tsg.pdf
SENSIT 14-day trial link, Terms & Conditions, Release Notes, and User Requirements can all be found on the SENSIT download page at:
www.futek/sensit-software-download SENSIT software keys may be looked up on the FUTEK support page under the Find my software key link.
www.futek/support/productkey
The SENSIT software manual and further support can be found on the FUTEK SENSIT support page: www.futek/sensit-software-support Profiles used for sensor system calibrations are listed on the calibration cert for instruments with multiple calibration profile capability.
Readings captured from a Data Logging or Live Graphing session are time stamped by the computer upon a successfulunable
read of packet information and are down to second units.
3
4
If available, calibration profiles in USB instruments can be selected in the drop-down menu found by right clicking in Display Mode.
Supply voltage options for amplified sensors from USB instruments are accessible from the drop-down menu in
Display Mode with a mouse right click.
Potential Error Messages
“Conversion from string “Error” to type ‘Integer’ is not Array valid.”
If the USB instrument is unable to generate an ADC number
from the sensor’s incoming signal, the “Error” notification will
be displayed instead. SENSIT will generate this error when it
receives the string word, “Error,” instead of an integer number.
Possible causes could be if the internal USB instrument
calibration was incorrect, and the instrument could not
generate an ADC number, or if there was an issue with the
signal from the sensor, such as noise or disconnection.
5