800 Dosino – 开始参数
标签: 方法 设备 Dosino
溶液
Select the solution for the Dosino that should be available when the hardware is started with [Start HW]. A check is carried out in the method run to ensure that the correct solution has been put on the dosing device. When the method is started, the working life and GLP test interval are checked for the selected solution.
选择
未定义 | Solutions from the solution table
默认值
未定义
未定义
The solution is not checked.
Solutions from the solution table
Only "intelligent" solutions, i.e. solutions that are assigned to an 807 Dosing unit, can be selected.
Connector
Shows the main device and the MSB connector which the 800 Dosino is on.
800 Dosino – 命令
Overview of the time program commands available for the 800 Dosino:
加液
对话窗口:方法 时间程序 编辑 ‘设备’ – 加液
The device-dependent time program command Dosing (command with feedback) doses the specified volume via the defined port. There is no automatic filling beforehand or afterwards.
参数
Port
Port via which dosing takes place. Formula input possible.
范围
1 ... 4
默认值
1
Volume
Volume to be transported. Formula input possible.
范围
0.0000 ... 99999.9000 mL
默认值
1.0000 mL
Dosing rate
Speed at which discharge takes place. The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
范围
0.01 ... 6.00 mL/min
默认值
1.00 mL/min
选择
最大值
Filling rate
Speed at which filling takes place. The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
范围
0.01 ... 6.00 mL/min
默认值
1.00 mL/min
选择
最大值
命令
时间程序命令的可选注释。
Input
64 characters
生成的变量
The following type 'TPmand number.Variable name' variables are generated by the command:
Variable name
Description
Data type
CONC
Concentration of the dosed solution as long as one has been assigned to the Dosino.
Number
DVOL
Dosed volume [mL]
Number
吸液
对话窗口:方法 时间轴 编辑 ‘设备’ – 吸液
The device-dependent time program command Aspirate (command with feedback) draws in the specified volume via the defined port. As with the command Dosing, the cylinder is not filled automatically beforehand or afterwards. It should be possible to achieve the volume to be aspirated with a single piston stroke.
参数
Port
Port via which aspiration takes place. Formula input possible.
input命令
范围
1 ... 4
默认值
1
Volume
Volume to be transported. Formula input possible.
范围
0.0000 ... 50.0000 mL
默认值
1.0000 mL
Rate
The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
范围
0.01 ... 6.00 mL/min
选择
最大值
默认值
最大值
命令
时间程序命令的可选注释。
Input
64 characters
生成的变量
The following type 'TPmand number.Variable name' variables are generated by the command:
Variable name
Description
Data type
AVOL
Aspirated volume [mL]
Number
充液
对话窗口:方法 时间轴 编辑 ‘设备’ – 充液
The device-dependent time program command Fill (command with feedback) fills the cylinder via the specified port. The valve disc stops on the selected port.
参数
Port
Port via which filling takes place. Formula input possible.
范围
1 ... 4
默认值
2
Rate
The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
范围
0.01 ... 6.00 mL/min
选择
最大值
默认值
最大值
命令
时间程序命令的可选注释。
Input
64 characters
排空
对话窗口:方法 时间轴 编辑 ‘设备’ – 排空
The device-dependent time program command Eject to end volume (command with feedback) ejects the entire contents of the cylinder via the specified port. In contrast to the command Eject to stop, the piston travels to the 最大值 volume until it has performed 10,000 pulses. This command should be used for pipetting functions for emptying the cylinder.
参数
Port
Port via which ejection takes place. Formula input possible.
范围
1 ... 4
默认值
4
Rate
The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL
范围
0.01 ... 6.00 mL/min
选择
最大值
默认值
最大值
命令

版权声明:本站内容均来自互联网,仅供演示用,请勿用于商业和其他非法用途。如果侵犯了您的权益请与我们联系QQ:729038198,我们将在24小时内删除。