800 Dosino – 开始参数
标签: 方法 ▶ 设备 ▶ Dosino
溶液
Select the solution for the Dosino that should be available when the hardware is started with [Start HW]. A check is carried out in the method run to ensure that the correct solution has been put on the dosing device. When the method is started, the working life and GLP test interval are checked for the selected solution.
选择 | 未定义 | Solutions from the solution table |
默认值 | 未定义 |
未定义
The solution is not checked.
Solutions from the solution table
Only "intelligent" solutions, i.e. solutions that are assigned to an 807 Dosing unit, can be selected.
Connector
Shows the main device and the MSB connector which the 800 Dosino is on.
800 Dosino – 命令
Overview of the time program commands available for the 800 Dosino:
加液
对话窗口:
The device-dependent time program command Dosing (command with feedback) doses the specified volume via the defined port. There is no automatic filling beforehand or afterwards.
参数
Port
Port via which dosing takes place. Formula input possible.
范围 | 1 ... 4 |
默认值 | 1 |
Volume
Volume to be transported. Formula input possible.
范围 | 0.0000 ... 99999.9000 mL |
默认值 | 1.0000 mL |
Dosing rate
Speed at which discharge takes place. The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL | |
范围 | 0.01 ... 6.00 mL/min |
默认值 | 1.00 mL/min |
选择 | 最大值 |
Filling rate
Speed at which filling takes place. The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL | |
范围 | 0.01 ... 6.00 mL/min |
默认值 | 1.00 mL/min |
选择 | 最大值 |
命令
时间程序命令的可选注释。
Input | 64 characters |
生成的变量
The following type 'TPmand number.Variable name' variables are generated by the command:
Variable name | Description | Data type |
CONC | Concentration of the dosed solution as long as one has been assigned to the Dosino. | Number |
DVOL | Dosed volume [mL] | Number |
吸液
对话窗口:
The device-dependent time program command Aspirate (command with feedback) draws in the specified volume via the defined port. As with the command Dosing, the cylinder is not filled automatically beforehand or afterwards. It should be possible to achieve the volume to be aspirated with a single piston stroke.
参数
Port
Port via which aspiration takes place. Formula input possible.
input命令范围 | 1 ... 4 |
默认值 | 1 |
Volume
Volume to be transported. Formula input possible.
范围 | 0.0000 ... 50.0000 mL |
默认值 | 1.0000 mL |
Rate
The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL | |
范围 | 0.01 ... 6.00 mL/min |
选择 | 最大值 |
默认值 | 最大值 |
命令
时间程序命令的可选注释。
Input | 64 characters |
生成的变量
The following type 'TPmand number.Variable name' variables are generated by the command:
Variable name | Description | Data type |
AVOL | Aspirated volume [mL] | Number |
充液
对话窗口:
The device-dependent time program command Fill (command with feedback) fills the cylinder via the specified port. The valve disc stops on the selected port.
参数
Port
Port via which filling takes place. Formula input possible.
范围 | 1 ... 4 |
默认值 | 2 |
Rate
The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL | |
范围 | 0.01 ... 6.00 mL/min |
选择 | 最大值 |
默认值 | 最大值 |
命令
时间程序命令的可选注释。
Input | 64 characters |
排空
对话窗口:
The device-dependent time program command Eject to end volume (command with feedback) ejects the entire contents of the cylinder via the specified port. In contrast to the command Eject to stop, the piston travels to the 最大值 volume until it has performed 10,000 pulses. This command should be used for pipetting functions for emptying the cylinder.
参数
Port
Port via which ejection takes place. Formula input possible.
范围 | 1 ... 4 |
默认值 | 4 |
Rate
The 最大值 rate depends on the cylinder volume of the buret used. Formula input possible.
Cylinder volume 2 mL | |
范围 | 0.01 ... 6.00 mL/min |
选择 | 最大值 |
默认值 | 最大值 |
命令